High Sensitivity&TE-Cooled Local:Home > ATP5020

ATP5020

◆Specifications:

Wavelength(nm)ModelOutput power (mW)Data sheets

ATP5020 UV-Vis Spectrometer employs the ultra-high performance, high-sensitivity, high resolution, and cost-effective model. The award-winner spectrometer is called cooled Maya, it can even replace Maya and QE65 in SNR, sensitivity, and resolution. It's ideal for Raman system well. It uses 2048 x 64 pixel, a back-thinned CCD array, TE-cooled down to -10 ℃. Sensor noise is reduced, and SNR is almost 2 times higher than others. Increased reliability can remain measure result do not change with temperature.
Customized design of ultra-low noise CCD signal processing circuit built inside, Noise is lower can 3 counts is ahead of records in the industries. 
ATP5020 is designed with SMA905 optic fiber entrance connector or free space, and it outputs spectral data via USB2.0/UART
DETECTOR
Detector Type Back-thinned Linear array detector, TE-cooled to -10℃
Spectral Range 180-1180 nm
Effective Pixels 2048
Pixels Dimension 14μm×14μm
Full Well Capacity 200 Ke-
Detector Model Hamamtsu s10420+TEC or s11850
Sensitivity 6.5 uV/e-
Dark Noise 300 e-rms
OPTICAL PARAMETER
Wavelength Range 180-1180nm
Optical Resolution 0.06-2 nm
Signal-To-Noise >1300:1
Dynamic Range 50000
Working Temperature -10°c to + 50°c
Numberical Aperture 0.11
Working Humidity < 90% NC
Trigger YES (IN and OUT)
Cooling Temperature -5°c
Full Well 300 ke-
Scanning Rate < 1 kfps
OPTICAL CONFIGURATION
Optical Design f/4 crossed asymmetrical Czerny-Turner
Focal Distance 40/60mm
Incident Slit 50 μm (5, 10, 25, 50, 100, or 200 μm optional)
Incident Interface SMA905 connector,free space
ELECTRICAL PARAMETER
Integration Time 1 ms - 30 s
Interfaces USB 2.0 (480MHz High speed)
A/D Conversion Resolution 18 bit
Supply Voltage DC4.5 to 5.5 V (type @5V)
Operating Current <2.3A
Storage Temperature -30°c to + 70 °c
PHYSICAL PARAMETERS
Dimension 170×110×52 mm^3
Weight 0.8kg
Sealing Anit-sweat
Feature:
Detector:back-thinned CCD, TE-cooled down to -10 ℃
CCD parameters:2048×64pixel, 14×14um
Ultra-low noise CCD signal processing circuit
Spectral range: 185-1100nm
Spectral resolution: 0.1-2 nm(depend on spectral range, slit width)
Optical path:crossed Czerny-Turner (C-T)
Integration time:2ms-130s
power supply:DC 5V±10% @ <2.3A
18 bit, 570KHz A/D Converter
Entrance connector:SMA905 connector or free space
Output interface: high-speed USB2.0 or UART
20 pins, dual rows programmable extension connector